ISO 8655 Defined

ISO 8655 Procedures

ISO 8655:2002, was written specifically to define the requirements necessary to produce accurate and reliable calibrations of piston pipettes and other closely related measurement equipment. It details the required methods, test conditions, test equipment, reporting requirements and includes requirements for reporting the required measurement uncertainty values. Today, it is the most critical ISO standard for calibrating piston-operated pipettes, burettes, diluters, and dispensers.

 

How is it different than ISO 17025?

ISO 8655, on the other hand, is a very detailed set of instructions, methods and controls that a pipette calibration provider must follow to correctly and accurately calibrate piston pipettes. The difference will explain how, for example, a traveling or onsite pipette calibration service can be accredited to ISO 17025 without following all instructions defined in ISO 8655, leading to wider measurement uncertainties and variations in measurement integrity. For more on the differences between the two ISO standards, see Support for ISO 8655.

 

ISO 8655 Defined

Adherence to the ISO 8655 standard covers air displacement pipettes and positive displacement pipettes, single channel and multichannel pipettes (both fixed and adjustable volume) as well as both manual pipettes and automated pipettes, often called motorized pipettes or electronic pipettes. Let’s look deeper into the quality that is ISO 8655.

The ISO 8655 standard precisely defines 6 elements required for accurate, repeatable, ISO-compliant measurements:

Acceptable Measurement Uncertainties

Measurement uncertainty is defined as a quantitative value representing a calculated level of ‘doubt’ or potential measurement error occurring from external influences, and based on the quality, standards and controls maintained or omitted from a calibration event. ISO accredited laboratories like Transcat are required to verify, calculate and report measurement uncertainty for every calibration.

 

 

 

Maximum Permissible Errors

Maximum Error Limits for piston-operated pipettes are clearly defined in ISO 8655. The standard characterizes both the maximum permissible systematic error, as well as the maximum permissible random error limits for a device at specific volumes ranging from 1-10,000uL. These errors are doubled for multichannel pipettes. Most pipette manufacturer’s tolerance specifications fall well below these limits, but at low volume measurements, meeting ISO 8655 tolerances can be a challenge for many service providers.

Maximum permissible errors for Air Displacement and Certain Positive Displacement pipettes – [Reference: ISO8655-2:2002, pg. 6]

Table 1 - Maximum Permissible Errors for Types A and D1

Nominal Volume

Maximum Permissible Systematic Error

Maximum Permissible Random Error

µl

±%

±µla

±%b

±µlc

1

5,0

0,05

5,0

0,05

2

4,0

0,08

2,0

0,04

5

2,5

0,125

1,5

0,075

10

1,2

0,12

0,8

0,08

20

1,0

0,2

0,5

0,1

50

1,0

0,5

0,4

0,2

100

0,8

0,8

0,3d

0,3d

200

0,8

1,6

0,3d

0,6d

500

0,8

4,0

0,3

1,5

1 000

0,8

0,8

0,3

3,0

2 000

0,8

16

0,3

6,0

5 000

0,8

40

0,3

15,0

10 000

0,6

60

0,3

30,0

a Expressed as the deviation of the mean of a tenfold measurement from the nominal or selected volume (see ISO 8655-6:2002, 8.4).

b Expressed as the coefficient of variation of a tenfold measurement (see ISO 8655-6:2002, 8.5).

c Expressed as the repeatability standard deviation of a tenfold measurement (see ISO 8655-6:2002, 8.4).

d For piston pipettes of type D1 the maximum permissible errors may be ±0,4%

 

 

Table 2 - Maximum Permissible Errors for Type D2

Nominal Volume

Maximum Permissible Systematic Error

Maximum Permissible Random Error

µl

±%

±µla

±%b

±µlc

5

2,5

0,13

1,5

0,08

10

2,0

0,2

1,0

0,1

20

2,0

0,4

0,8

0,16

50

1,4

0,7

0,6

0,3

100

1,5

1,5

0,6

0,3

200

1,5

3,0

0,4

0,8

500

1,2

6,0

0,4

2,0

1 00

1,2

12,0

0,4

4,0

a Expressed as the deviation of the mean of a tenfold measurement from the nominal or selected volume (see ISO 8655-6:2002, 8.4).

b Expressed as the coefficient of variation of a tenfold measurement (see ISO 8655-6:2002, 8.5).

c Expressed as the repeatability standard deviation of a tenfold measurement (see ISO 8655-6:2002, 8.4).

 

Methodology

ISO 8655 states that the primary methodology for measurement of piston-operated pipettes and related measurement apparatus is through gravimetric measurement analysis. Included are the minimum balance requirements at each test volume, important for reporting measurement accuracy at the appropriate resolution. For example, you wouldn’t weigh yourself on a truck scale. Pipette measurements face the same challenge for many providers. This standard also requires control (not just monitoring) of test conditions, along with minimum data points for a valid representation of device performance and measurement analysis.

Minimum Balance Requirements – [Reference: ISO8655-6:2002, pg. 2]

 

Lab and Environmental Conditions

To be considered valid by ISO 8655, pipette measurements must occur in a strictly controlled, vibration-free test environment. The acceptable ranges of environmental conditions are defined as:

  • Temperature – constant [±0.5°C], 15-30°C
  • Relative Humidity – ˃50%
  • Air Flow – draft free
  • Static – ~0
  • Evaporation Rate – ~0
  • Vibration – ~0
  • Altitude – Ground Level

Process Requirements

The ISO 8655 standard defines a step-wise procedure for generating valid, reliable measurement data. Procedures include proper technique for tip installation, pre-wetting, aspiration and dispensing technique, measurement container requirements, required number of measurements, evaporation rate determination and time lapse for test completion.

ISO 8655 and FDA Audits

Regular FDA audits are required for all drug development labs, helping to ensure data integrity and proof of efficacy. However, supporting your science with the assurance of an ISO 8655 compliant pipette calibration service provider, ensures that the pipette measurements are accurate and reliable. As FDA audits validate proper drug development and quality controls, ISO 17025 accreditation audits provide the same integrity for pipette calibration providers. However, when an auditor evaluates compliance of a pipette service provider to ISO 8655 standards and controls, they are raising the bar and promoting standards-based calibration. For pipette service providers, ISO 8655 compliance improves service integrity, so drug development labs provide quality products, supported by trustworthy, reliable service providers.

To see how we can help you with your pipette calibration services, contact us at 1-800-242-6022 or email us at pipetteinfo@transcat.com.

or give us a call at 1-800-242-6022 for additional information.